-counters in the Mass-8 experiment
J.P.S. van Schagen, K.B. Swartz and D. Wright
To obtain the coefficients a1 and a2 in the
angular correlation (see
Section 1.1) as measured with the Mass-8 apparatus, an accurate
knowledge of the calibration of
the
-counters is necessary.
This calibration must be obtained in situ, since no natural
mono-chromatic
-sources of sufficient energy are available.
However, the
-decay feeds a broad final state in
8Be which subsequently decays
into two
-particles.2
A Si-counter placed at 90° with respect
to all seven
-counters measured the
-particle kinetic energy
T
with high resolution. This energy, corrected for the energy loss
in the catcher foil, can be related to the (total relativistic)
-endpoint energy E0 by:
(1)
In measuring
-particles in coincidence with an
-particle in the
Si-counter, spectra for different endpoint energies were obtained.
These spectra were then fitted simultaneously to the theoretical
spectral distribution for the electron/positron decay in a
-minimization
procedure. Only the values for the offset
acal,
the linear coefficient bcal
and the normalization constant
Nnrom for each value E0 were varied.
The spectral distribution is given by:
(2)
which is valid for allowed transitions and includes folding with the
-counter response.
F(Z,E'
,
+/-)
is the Fermi-function, pc the
electron/positron momentum. For the lineshape
R(E
,E'
)
a Gaussian peak
with an exponential tail is used.3
As a typical example, Fig. 1.2-1 shows the results of the fit for an endpoint energy E0 = 14.262 MeV where both the data and the fit have been converted to a Kurie-plot for easy comparison. The data in between the two markers were included in the region. Excellent agreement can be observed.
Fig. 1.2-1. Data and theoretical spectral distribution plotted as a Kurie-plot for E0 = 14.3 MeV. The calibration coefficients are given by acal = 1.43 MeV, bcal = 0.0135 MeV/CH.